RecallRadar

FDA Device recall Z-0190-2027

The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists

On 2026-10-21 the FDA classified a Class II recall of The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists by Radiometer Medical Aps, citing in vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer s….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0190-2027
ClassificationClass II
Statusongoing
Initiated2026-08-18
Published2026-10-21
Last updated2026-10-27 05:30 UTC
CompanyRadiometer Medical Aps
DistributionUS

Product

The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order.

Reason

In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings.

Identifiers

code_info
UDI-DI: 05700693930923, 05700693930909 Software versions 3.6 and lower. ABL90 FLEX PLUS Instructions for Use versions p…UDI-DI: 05700693930923, 05700693930909 Software versions 3.6 and lower. ABL90 FLEX PLUS Instructions for Use versions prior to: 996-223, Version: 202606O. ABL90 FLEX Instructions for Use versions prior to: 996-189, Version: 202606L

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-10-27 05:30 UTC. Current status: ongoing.

Other recalls by Radiometer Medical Aps

All Radiometer Medical Aps recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0190-2027%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.