RecallRadar

FDA Device recall Z-0187-2021

7" (18cm) Appx 0.31 ml, Smallbore Ext Set w/Remv MicroClave¿ Clear, NanoClave T-Connector, Clamp, Rotating Luer. 1 unit per pouch, 50 multi…

On 2020-10-28 the FDA classified a Class II recall of 7" (18cm) Appx 0.31 ml, Smallbore Ext Set w/Remv MicroClave¿ Clear, NanoClave T-Connector, Clamp, Rotating Luer. 1 unit per pouch, 50 multi… by Icu Medical, citing identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoCl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0187-2021
ClassificationClass II
Statusterminated
Initiated2020-07-06
Published2020-10-28
Last updated2026-09-13 11:55 UTC
CompanyIcu Medical
DistributionUS

Product

7" (18cm) Appx 0.31 ml, Smallbore Ext Set w/Remv MicroClave¿ Clear, NanoClave T-Connector, Clamp, Rotating Luer. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709051055(17)250101(30)50(10)4590834; (01)10887709051055(17)250101(30)50(10)4557481; (01)10887709051055(17)241201(30)50(10)4536601; (01)10887709051055(17)250201(30)50(10)4724818; (01)10887709051055(17)250201(30)50(10)4733548. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).

Reason

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Identifiers

code_infoLot Numbers:4590834, 4557481, 4536601, 4724818, 4733548.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0187-2021%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.