FDA Device recall Z-0186-2020
Torrent Suite Dx Software Version (IUO) 5.4 Catalog Number: A31774 OEM Customer Software: Sentosa SQ Suite Software v5.6.15
- FDA Device
- Class II
- terminated
- Published 2019-10-30
On 2019-10-30 the FDA classified a Class II recall of Torrent Suite Dx Software Version (IUO) 5.4 Catalog Number: A31774 OEM Customer Software: Sentosa SQ Suite Software v5.6.15 by Life Technologies Holdings Pte, citing torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is det….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0186-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-09-26 |
| Published | 2019-10-30 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Life Technologies Holdings Pte |
| Distribution | AZ, CA, CO, CT, FL, GA, IN, MA, MD, NC, NJ, NM, NY, OR, TX, VA, WA |
Product
Torrent Suite Dx Software Version (IUO) 5.4 Catalog Number: A31774 OEM Customer Software: Sentosa SQ Suite Software v5.6.15
Reason
Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant
Identifiers
| code_info | The field action is associated with the software available on the PGM Dx instrument, therefore no lot code is applicable for this part type |
|---|