FDA Device recall Z-0183-2026
Medical convenience kits Item Number/Description HSLM53AB SPINE PACK MILM82A DTC LAMINECTOMY PACK (PS 209475) RCNA47B NEURO ANGIO 0246971 S…
- FDA Device
- Class II
- ongoing
- Published 2025-10-22
On 2025-10-22 the FDA classified a Class II recall of Medical convenience kits Item Number/Description HSLM53AB SPINE PACK MILM82A DTC LAMINECTOMY PACK (PS 209475) RCNA47B NEURO ANGIO 0246971 S… by American Contract Systems, citing identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have no….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0183-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-09-04 |
| Published | 2025-10-22 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | American Contract Systems |
| Distribution | US |
Product
Medical convenience kits Item Number/Description HSLM53AB SPINE PACK MILM82A DTC LAMINECTOMY PACK (PS 209475) RCNA47B NEURO ANGIO 0246971 SFLM54AB LAMINECTOMY PACK - 219659 UDLM94AA LAMINECTOMY PACK - 206064 UILM77AU LAMINECTOMY (PS 039500) UINR44P NEURO IR PERIPHERAL (PS 64405) UISH15AV SHUNT PACK (PS 039360)
Reason
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Identifiers
| code_info | ANLM40AG UDI-DI 191072232076 Lot 8192111 Bag serial number 68195596; ANPP15AR UDI-DI 191072214423 Lot 8024911 Bag seri…ANLM40AG UDI-DI 191072232076 Lot 8192111 Bag serial number 68195596; ANPP15AR UDI-DI 191072214423 Lot 8024911 Bag serial number 68055493; HSLM53AB UDI-DI 191072224095 Lot 8061111 Bag serial number 68138767; MILM82A UDI-DI 191072231512 Lot 8361111 Bag serial number 68576542; RCNA47B UDI-DI 191072228536 Lot 8441711 Bag serial number 68680026; SFLM54AB UDI-DI 191072217400 lot 8105511 Bag serial number N/A; UDLM94AA UDI-DI 191072214089 Lot 8382011 Bag serial number 17621919; UILM77AU UDI-DI 191072210333 Lot 8056311 Bag serial number N/A; UINR44P UDI-DI 191072210340 Lot 8067511 Bag serial numbers 68138618 68140374; UISH15AV UDI-DI 191072227843 Lot 8028211 Bag serial number 68154067 * If bag serial number is listed as "N/A", then whole lot is affected |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0183-2026%22