Applied Medical's Inzii¿ 12/15mm Retrieval System

On 2013-11-20 the FDA classified a Class II recall of Applied Medical's Inzii¿ 12/15mm Retrieval System by Applied Medical Resources, citing applied Medical is conducting a voluntary recall of the Inzii¿ 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0183-2014
ClassificationClass II
Statusterminated
Initiated2013-10-21
Published2013-11-20
Last updated2026-09-13 11:53 UTC
CompanyApplied Medical Resources
DistributionUS

Product

Applied Medical's Inzii¿ 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is classified as a Class II device. It was cleared to market via 510(k) filing K060051. The retrieval bag is a disposable specimen retrieval bag intended for use as receptacle for the collection of tissue, organs, and calculi during laparoscopic procedures. Applieds CD004 has a shelf life of three years. It is provided sterile and packaged in a heat sealed Tyvek pouch.

Reason

Applied Medical is conducting a voluntary recall of the Inzii¿ 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with small holes, which could compromise the sterile barrier. The likelihood of this situation to occur and result in patient harm is highly unlikely; however, out of an abundance of caution for patient safety and a commitmen

Identifiers

code_info
Model Number CD004. Lot Nubmers: 1161268, 1161269, 1164934, 1168358, 1168361, 1169317, 1169318, 1170694, 1170700, 1170…Model Number CD004. Lot Nubmers: 1161268, 1161269, 1164934, 1168358, 1168361, 1169317, 1169318, 1170694, 1170700, 1170701, 1171214, 1172482, 1172664, 1174287, 1174793, 1174857, 1175123, 1175280, 1175472, 1176555, 1177023, 1177871, 1179069, 1179392, 1179460, 1179775, 1179890, 1180291, 1180411, 1180576, 1181647, 1181648, 1181649, 1181939, 1182760, 1182934, 1183225, 1183823, 1183957, 1184273, 1184728, 1184915, 1184916, 1187779, 1188817, 1189117, 1189957, 1190294, 1190358, 1191208, 1192057, 1192969, 1193196, 1193661, 1194010, 1194666, 1194807, 1195518, 1195727, 1195997, 1197182, 1197404, 1197832, 1198055, 1198968, 1199899, 1200658, 1200828, 1200829, 1200830, 1201739, 1201740, 1201741, 1203154, 1204054.