RecallRadar

FDA Device recall Z-0182-2026

Medical convenience kits Item Number/Description LKRA40G RADIOLOGY PACK

On 2025-10-22 the FDA classified a Class II recall of Medical convenience kits Item Number/Description LKRA40G RADIOLOGY PACK by American Contract Systems, citing identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have no….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0182-2026
ClassificationClass II
Statusongoing
Initiated2025-09-04
Published2025-10-22
Last updated2026-09-13 11:57 UTC
CompanyAmerican Contract Systems
DistributionUS

Product

Medical convenience kits Item Number/Description LKRA40G RADIOLOGY PACK

Reason

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Identifiers

code_infoLKRA40G UDI-DI 191072189691 Lot 8025211 Bag serial number 68154347 * If bag serial number is listed as "N/A", then whole lot is affected

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0182-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.