FDA Device recall Z-0180-2015
Covidien Puritan Bennett 980 Ventilator, Rx ONLY, Made in Ireland., Covidien LLC, USA. REF 980X1ENDIUU
- FDA Device
- Class I
- terminated
- Published 2014-12-17
On 2014-12-17 the FDA classified a Class I recall of Covidien Puritan Bennett 980 Ventilator, Rx ONLY, Made in Ireland., Covidien LLC, USA. REF 980X1ENDIUU by Nellcor Puritan Bennett Dba Covidien, citing covidien is recalling certain Puritan Bennett 980 Ventilator Systems due to a blacklight inverter PCBA where a capacitor component can fail, causing a dim but functional screen an….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0180-2015 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2014-10-01 |
| Published | 2014-12-17 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Nellcor Puritan Bennett Dba Covidien |
| Distribution | US |
Product
Covidien Puritan Bennett 980 Ventilator, Rx ONLY, Made in Ireland., Covidien LLC, USA. REF 980X1ENDIUU.
Reason
Covidien is recalling certain Puritan Bennett 980 Ventilator Systems due to a blacklight inverter PCBA where a capacitor component can fail, causing a dim but functional screen and thermal odor.
Identifiers
| code_info | Model number PB980 Ventilator:35B1400143, 35B1400684, 35B1400627, 35B1400708, 35B1400398, 35B1400482, 35B1400491, 35B…Model number PB980 Ventilator:35B1400143, 35B1400684, 35B1400627, 35B1400708, 35B1400398, 35B1400482, 35B1400491, 35B1400512, 35B1400601, 35B1400121, 35B1400568, 35B1400590, 35B1400595, 35B1400599, 35B1400606, 35B1400106, 35B1400146, 35B1400169, 35B1400226, 35B1400260, 35B1400270, 35B1400263, 35B1400351, 35B1400373, 35B1400400, 35B1400401, 35B1400608, 35B1400089, 35B1400112, 35B1400683, 35B1400685, 35B1400678, 35B1400118, 35B1400114, 35B1400355, 35B1400144, 35B1400300, 35B1400379, 35B1400412, 35B1400472, 35B1400489, 35B1400490, 35B1400492, 35B1400493, 35B1400499, 35B1400500, 35B1400501, 35B1400502, 35B1400507, 35B1400577, 35B1400207, 35B1400211, 35B1400216, 35B1400222, 35B1400225, 35B1400231, 35B1400267, 35B1400312, 35B1400323, 35B1400340, 35B1400383, 35B1400128, 35B1400363, 35B1400459, 35B1400589, 35B1400616, 35B1400637, 35B1400483, 35B1400311, 35B1400447, 35B1400648, 35B1400651, 35B1400654, 35B1400662, 35B1400686, 35B1400692, 35B1400175, 35B1400294, 35B1400436, 35B1400485, |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0180-2015%22