FDA Device recall Z-0178-2026
Medical convenience kits Item Number/Description RCEN25B ENT SDSC 0246972 RCHN27D HEAD AND NECK RCH 0246951
- FDA Device
- Class II
- ongoing
- Published 2025-10-22
On 2025-10-22 the FDA classified a Class II recall of Medical convenience kits Item Number/Description RCEN25B ENT SDSC 0246972 RCHN27D HEAD AND NECK RCH 0246951 by American Contract Systems, citing identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have no….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0178-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-09-04 |
| Published | 2025-10-22 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | American Contract Systems |
| Distribution | US |
Product
Medical convenience kits Item Number/Description RCEN25B ENT SDSC 0246972 RCHN27D HEAD AND NECK RCH 0246951
Reason
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Identifiers
| code_info | RCEN25B UDI-DI 191072225115 lot 8138311 bag serial number N/A; RCHN27D UDI-DI 191072235923 lot 8508211 bag serial numb…RCEN25B UDI-DI 191072225115 lot 8138311 bag serial number N/A; RCHN27D UDI-DI 191072235923 lot 8508211 bag serial number 47847459 * If bag serial number is listed as "n/a", then entire lot is affected |
|---|