RecallRadar

FDA Device recall Z-0178-2023

Beach Chair Clamp, BEACH CHAIR SHLDER POSITIONER CMPT, REF A-90013, #800104

On 2022-11-09 the FDA classified a Class II recall of Beach Chair Clamp, BEACH CHAIR SHLDER POSITIONER CMPT, REF A-90013, #800104 by Baxter Healthcare, citing the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integri….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0178-2023
ClassificationClass II
Statusterminated
Initiated2022-09-20
Published2022-11-09
Last updated2026-09-13 11:56 UTC
CompanyBaxter Healthcare
DistributionMI, TX, WA, WI

Product

Beach Chair Clamp, BEACH CHAIR SHLDER POSITIONER CMPT, REF A-90013, #800104. Component to position and support the patients head in surgical procedures.

Reason

the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.

Identifiers

code_infoUDI/DI 00615521003920, Lot Number: 10-22-140

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0178-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.