FDA Device recall Z-0176-2022
ASC PneumoLiner device, Part No. WA90500US (US market only) The Pneumoliner device is intended for use as a multiple instrument port and ti…
- FDA Device
- Class II
- terminated
- Published 2021-11-03
On 2021-11-03 the FDA classified a Class II recall of ASC PneumoLiner device, Part No. WA90500US (US market only) The Pneumoliner device is intended for use as a multiple instrument port and ti… by Olympus, citing due to manufacturing error, the Pneumoliner Bag Distal Tab that exits the Introducer shaft is in the wrong orientation. This results in the user deploying the bag upside down, whi….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0176-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-09-21 |
| Published | 2021-11-03 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Olympus |
| Distribution | US |
Product
ASC PneumoLiner device, Part No. WA90500US (US market only) The Pneumoliner device is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue, considered benign, resected during single-port or multi- site laparoscopic surgery during power morcellation and removal.
Reason
Due to manufacturing error, the Pneumoliner Bag Distal Tab that exits the Introducer shaft is in the wrong orientation. This results in the user deploying the bag upside down, which will make tissue encapsulation and bag closure more difficult, introducing the risk of trapping the small bowel/viscera in the bag at closure resulting in patient injury.
Identifiers
| code_info | Lots 647572 and 667060 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0176-2022%22