FDA Device recall Z-0173-2020

Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01

On 2019-10-30 the FDA classified a Class II recall of Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01 by Abbott Laboratories, citing the diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner and outer surfaces which may result in falsely elevated platelet results.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0173-2020
ClassificationClass II
Statusterminated
Initiated2019-05-09
Published2019-10-30
Last updated2026-09-13 11:55 UTC
CompanyAbbott Laboratories
DistributionGA, LA, MA, NC, NY, SC, TX

Product

Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL.

Reason

The diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner and outer surfaces which may result in falsely elevated platelet results.

Identifiers

code_info
Tubing lot #2018-10-16; Instrument serial numbers 007689, 007690, 007691, 007692, 007705, 007706, 007707, 007708, 00770…Tubing lot #2018-10-16; Instrument serial numbers 007689, 007690, 007691, 007692, 007705, 007706, 007707, 007708, 007709, 007728, 007694, 007695, 007697, 007766, 007767, 007768, 007765, 007710, 007699, 007700, 007701, and 007703.