RecallRadar

FDA Device recall Z-0170-2024

Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog…

On 2023-11-08 the FDA classified a Class I recall of Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog… by Asensus Surgical, citing asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparo….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0170-2024
ClassificationClass I
Statusterminated
Initiated2023-09-15
Published2023-11-08
Last updated2026-09-13 11:56 UTC
CompanyAsensus Surgical
DistributionFL, IL, LA, MN, NJ

Product

Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system

Reason

Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.

Identifiers

code_info
UDI/DI 00815440022075, Serial Numbers: SN 00012, SN 00016, SN 00007, SN 00017, SN 00035, SN 00018, SN 00042, SN 00008,…UDI/DI 00815440022075, Serial Numbers: SN 00012, SN 00016, SN 00007, SN 00017, SN 00035, SN 00018, SN 00042, SN 00008, SN 00031, SN 00025, SN 00011, SN 00028, SN 00038, SN 00029, SN 00036, SN 00025, SN 00024, SN 00006, SN 00019, SN 00040, SN 00041, loaded with Software (SW) version 2.7.4

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0170-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.