RecallRadar

FDA Device recall Z-0170-2023

WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System

On 2022-11-09 the FDA classified a Class II recall of WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System by Microvention, citing detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inser….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0170-2023
ClassificationClass II
Statusongoing
Initiated2022-09-21
Published2022-11-09
Last updated2026-09-13 11:56 UTC
CompanyMicrovention
DistributionUS

Product

WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.

Reason

Detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inserted into the controller, which may result in the inability of the controller to properly detach the implant device, which could cause delays and excessive manipulation that may cause the implant to move and damage vessels

Identifiers

code_info
REF/UDI/Lots: WDC-2/00842429103128/0000095683, 0000096284, 0000205967, 0000206629, 0000215962, 0000216862, 0000224074,…REF/UDI/Lots: WDC-2/00842429103128/0000095683, 0000096284, 0000205967, 0000206629, 0000215962, 0000216862, 0000224074, 0000241662, 0000242343, 0000242345; Distributed in China: WDC/00842429113684/0000078840, 0000086571, 0000088347, 0000088349, 0000136477, 0000181460, 0000191147; Distribution in Japan: MV-WDC2/04987892122231/0000142172; 0000166445; 0000210109; 0000245337

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0170-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.