RecallRadar

FDA Device recall Z-0167-2013

SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954

On 2012-11-07 the FDA classified a Class II recall of SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954 by Sorin Group Usa, citing sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0167-2013
ClassificationClass II
Statusterminated
Initiated2012-09-20
Published2012-11-07
Last updated2026-09-13 11:53 UTC
CompanySorin Group Usa
DistributionUS

Product

SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954. The SCP is a cardiopulmonary bypass speed control device indicated for use exclusively with the COBE Revolution pump head for speed-controlled pumping through cardiopulmonary bypass circuit for typical durations of six hours or less.

Reason

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.

Identifiers

code_infoProduct Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0167-2013%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.