RecallRadar

FDA Device recall Z-0159-2027

MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101

On 2026-10-21 the FDA classified a Class II recall of MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101 by Microgenics, citing vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which ma….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0159-2027
ClassificationClass II
Statusongoing
Initiated2026-08-17
Published2026-10-21
Last updated2026-10-27 05:30 UTC
CompanyMicrogenics
DistributionUS

Product

MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated Immunoassay Control, Sample Pack, REF: OPRO-SP

Reason

Vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.

Identifiers

code_infoREF/UDI-DI/Lot(Expiration): OPRO-101/00884883005327/OPRO29061A(2029-06-30); OPRO-SP/00884883005358/OPRO2906SA(2029-06-30)

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-10-27 05:30 UTC. Current status: ongoing.

Other recalls by Microgenics

All Microgenics recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0159-2027%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.