FDA Device recall Z-0159-2016
CC cartridges for IRMA TRUPOINT Blood Gas Analyzer, PN 039903
- FDA Device
- Class II
- terminated
- Published 2015-11-04
On 2015-11-04 the FDA classified a Class II recall of CC cartridges for IRMA TRUPOINT Blood Gas Analyzer, PN 039903 by Lifehealth, citing may report values outside the published performance specifications for pH. After a certain point in shelf life, the pH sensor may give results that are falsely elevated. The discr….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0159-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-09-16 |
| Published | 2015-11-04 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Lifehealth |
| Distribution | US |
Product
CC cartridges for IRMA TRUPOINT Blood Gas Analyzer, PN 039903. The CC Cartridge are single use, disposable, in vitro diagnostics intended for point of care (POC), professional, use with the IRMA TRUpoint¿ Blood Analysis System.
Reason
May report values outside the published performance specifications for pH. After a certain point in shelf life, the pH sensor may give results that are falsely elevated. The discrepancy is greater as shelf life progresses and can be seen on the aqueous controls before the blood.
Identifiers
| code_info | ASBEZ ASCXE ASEMB ASHWQ ASJMY ASLGI ASMXR ASOQO ASRXC ASTQR ASWZI |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0159-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.