RecallRadar

FDA Device recall Z-0157-2025

Medline brand medical procedure convenience kits, labeled as: a) KIT LINE INSERTION, REF CVI4730A

On 2024-10-30 the FDA classified a Class II recall of Medline brand medical procedure convenience kits, labeled as: a) KIT LINE INSERTION, REF CVI4730A by Medline Industries Northfield, citing it has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0157-2025
ClassificationClass II
Statusongoing
Initiated2024-09-04
Published2024-10-30
Last updated2026-09-13 11:57 UTC
CompanyMedline Industries Northfield
DistributionUS

Product

Medline brand medical procedure convenience kits, labeled as: a) KIT LINE INSERTION, REF CVI4730A; b) KIT LINE INSERTION, REF CVI4730A; c) LINE INSERTION TRAY, REF DYNJ20803D; d) CENTRAL LINE PACK-LF, REF DYNJ28368; e) CENTRAL LINE PACK-LF, REF DYNJ28368; f) PICC PACK-LF, REF DYNJ28487; g) PICC PACK-LF, REF DYNJ28487H; h) CENTRAL LINE PACK, REF DYNJ30087C; i) CENTRAL LINE ADULT PACK, REF DYNJ35352; j) PICC PACK, REF DYNJ52255; k) PICC PROCEDURE PACK, REF DYNJ59067A; l) PICC PROCEDURE PACK, REF DYNJ59067A; m) DR SIRAGUSA VEIN PACK, REF DYNJ61247A; n) PICC TEAM PACK, REF DYNJ64542A; o) PICC LINE INSERTION PACK, REF DYNJ65519; p) CVL INSERTION KIT II PGYBK, REF DYNJ68112; q) CVL INSERTION KIT II PGYBK, REF DYNJ68112; r) PICC PACK, REF DYNJ69088A; s) PICC INSERTION TRAY, REF DYNJ80031; t) ENDOVENOUS ABLATION PACK, REF DYNJ80489A; u) CVR ENDOVENOUS PACK, REF DYNJ80643; v) RADIAL ARTERY PACK, REF DYNJ83938;

Reason

It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.

Identifiers

code_info
a) REF CVI4730A, UDI/DI 40195327562039 (each), 10195327562038 (case), Lot Numbers: 24ABK903; b) REF CVI4730A, UDI/DI…a) REF CVI4730A, UDI/DI 40195327562039 (each), 10195327562038 (case), Lot Numbers: 24ABK903; b) REF CVI4730A, UDI/DI 40195327562039 (each), 10195327562038 (case), Lot Numbers: 23LBN351; c) REF DYNJ20803D, UDI/DI 40889942553885 (each), 10889942553884 (case), Lot Numbers: 24ABE939; d) REF DYNJ28368, UDI/DI 40080196200161 (each), 10080196200160 (case), Lot Numbers: 24ABD995; e) REF DYNJ28368, UDI/DI UDI/DI 40080196200161 (each), 10080196200160 (case), Lot Numbers: 23KBE939; f) REF DYNJ28487, UDI/DI 40080196227021 (each), 10080196227020 (case), Lot Numbers: 24ABI626; g) REF DYNJ28487H, UDI/DI40080196227021 (each), 10080196227020 (case), Lot Numbers: 24ABI626; h) REF DYNJ30087C, UDI/DI 40889942754145 (each), 10889942754144 (case), Lot Numbers: 23JBS113; i) REF DYNJ35352, UDI/DI 40884389433080 (each), 10884389433089 (case), Lot Numbers: 23LBN154; j) REF DYNJ52255, UDI/DI 40889942253709 (each), 10889942253708 (case), Lot Numbers: 24ABC063; k) REF DYNJ59067A, UDI/DI 4019532

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0157-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.