FDA Device recall Z-0156-2013
Repeater Pump, Class II device used for facilitating repeatable drug dosage distribution and reconstitution in hospital pharmacies
- FDA Device
- Class II
- terminated
- Published 2012-10-31
On 2012-10-31 the FDA classified a Class II recall of Repeater Pump, Class II device used for facilitating repeatable drug dosage distribution and reconstitution in hospital pharmacies by Baxa, citing the firm is issuing a field correction to the operators manual included with the 099 Repeater Pump because the Technical Manual was distributed with the device as opposed to the O….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0156-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-10-03 |
| Published | 2012-10-31 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Baxa |
| Distribution | US |
Product
Repeater Pump, Class II device used for facilitating repeatable drug dosage distribution and reconstitution in hospital pharmacies. The device are Fluid Transfer Tube Sets, Sterile/EO, Std Volume Trifurcated, BAXA Corporation, Englewood, CO. The Repeater Pump is a device that provides a peristaltic pump driven fluid transfer that facilitates repeatable drug dosage distribution and reconstitution in hospital pharmacies. The device is for use with IV bags, syringes, elastomeric infusers, and other drug administration containers.
Reason
The firm is issuing a field correction to the operators manual included with the 099 Repeater Pump because the Technical Manual was distributed with the device as opposed to the Operator Manual and the instruction for use contradict each other with regards to multiple source container use.
Identifiers
| code_info | Repeater Pump model #s 095, 099, and 099R. |
|---|