FDA Device recall Z-0156-2013

Repeater Pump, Class II device used for facilitating repeatable drug dosage distribution and reconstitution in hospital pharmacies

On 2012-10-31 the FDA classified a Class II recall of Repeater Pump, Class II device used for facilitating repeatable drug dosage distribution and reconstitution in hospital pharmacies by Baxa, citing the firm is issuing a field correction to the operators manual included with the 099 Repeater Pump because the Technical Manual was distributed with the device as opposed to the O….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0156-2013
ClassificationClass II
Statusterminated
Initiated2012-10-03
Published2012-10-31
Last updated2026-09-13 11:53 UTC
CompanyBaxa
DistributionUS

Product

Repeater Pump, Class II device used for facilitating repeatable drug dosage distribution and reconstitution in hospital pharmacies. The device are Fluid Transfer Tube Sets, Sterile/EO, Std Volume Trifurcated, BAXA Corporation, Englewood, CO. The Repeater Pump is a device that provides a peristaltic pump driven fluid transfer that facilitates repeatable drug dosage distribution and reconstitution in hospital pharmacies. The device is for use with IV bags, syringes, elastomeric infusers, and other drug administration containers.

Reason

The firm is issuing a field correction to the operators manual included with the 099 Repeater Pump because the Technical Manual was distributed with the device as opposed to the Operator Manual and the instruction for use contradict each other with regards to multiple source container use.

Identifiers

code_infoRepeater Pump model #s 095, 099, and 099R.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.