FDA Device recall Z-0153-2016

OneTray Sealed Sterilization Containers Product Usage: A rigid sterilization container used to steam sterilize medical devices

On 2015-10-28 the FDA classified a Class II recall of OneTray Sealed Sterilization Containers Product Usage: A rigid sterilization container used to steam sterilize medical devices by Innovative Sterlization Technologies, citing the wrong sterilization time was on the label. The label lists Steam Gravity 10 - 34 minutes at 270 degrees F instead of Steam Gravity 17 - 34 minutes at 270 degrees F.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0153-2016
ClassificationClass II
Statusterminated
Initiated2014-02-03
Published2015-10-28
Last updated2026-09-13 11:53 UTC
CompanyInnovative Sterlization Technologies
DistributionUS

Product

OneTray Sealed Sterilization Containers Product Usage: A rigid sterilization container used to steam sterilize medical devices

Reason

The wrong sterilization time was on the label. The label lists Steam Gravity 10 - 34 minutes at 270 degrees F instead of Steam Gravity 17 - 34 minutes at 270 degrees F.

Identifiers

code_infoM2104 Lot # 044305, 044379, 044380 & 044418. M2408 Lot # 044296, 022304, 044381, 044420, 044421 & 044422.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.