LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Control Kit Catalog #: L016080109002 (EUA) The LumiraDx SARS-CoV-2 Quality Controls are int…

On 2022-11-02 the FDA classified a Class III recall of LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Control Kit Catalog #: L016080109002 (EUA) The LumiraDx SARS-CoV-2 Quality Controls are int… by Lumiradx, citing assigned vial lot numbers for the incorrect type of product. SARS-CoV-2 & Flu A/B lot numbers were incorrectly assigned to the positive and negative vial lot numbers for SARS-CoV-….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0135-2023
ClassificationClass III
Statusterminated
Initiated2022-09-23
Published2022-11-02
Last updated2026-09-13 11:56 UTC
CompanyLumiradx
DistributionUS

Product

LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Control Kit Catalog #: L016080109002 (EUA) The LumiraDx SARS-CoV-2 Quality Controls are intended for liquid quality control testing performed on the LumiraDx Instrument when used with the LumiraDx SARS-CoV-2 Ag Test Strip. The Quality Controls provide users with assurance that the device is performing within specification

Reason

Assigned vial lot numbers for the incorrect type of product. SARS-CoV-2 & Flu A/B lot numbers were incorrectly assigned to the positive and negative vial lot numbers for SARS-CoV-2 Ag LQCs within batch 2002204017, which should have been assigned SARS-CoV-2 Ag lot numbers resulting in an error message and preventing use of the LQC

Identifiers

code_infoLQC Kit: 2002204017 Positive Control Vial: 1000 1710 0014 0141 Negative Control Vial: 1000 1720 0014 0142 Exp. Date: 01-Mar-2023