FDA Device recall Z-0135-2015
HbA1c Calibrators LN 4P52-02 consists of: 4P52-02 Hemoglobin A1c 1 x 1.6 mL and 1 x 1.6 mL. A1c Calibrators (lyophilized) contain hemoglobi…
- FDA Device
- Class II
- terminated
- Published 2014-11-05
On 2014-11-05 the FDA classified a Class II recall of HbA1c Calibrators LN 4P52-02 consists of: 4P52-02 Hemoglobin A1c 1 x 1.6 mL and 1 x 1.6 mL. A1c Calibrators (lyophilized) contain hemoglobi… by Abbott Laboratories, citing hbA1c Calibrators LN 4P52-02, lot 45063UQ12 may contain a value sheet for a different HbA1c calibrator lot. If calibrator lot 45063UQ12 is used in the calibration of the HbA1c ass….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0135-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-04-30 |
| Published | 2014-11-05 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Abbott Laboratories |
| Distribution | n/a |
Product
HbA1c Calibrators LN 4P52-02 consists of: 4P52-02 Hemoglobin A1c 1 x 1.6 mL and 1 x 1.6 mL. A1c Calibrators (lyophilized) contain hemoglobin and glycated hemoglobin from human whole blood. Prior to lyophilization, the calibrator matrix is an MES-buffered solution. Preservative: Ofloxacin. For use in the calibration of the Hemoglobin A1c assay on the ARCHITECT c 8000 and c 4000 Systems.
Reason
HbA1c Calibrators LN 4P52-02, lot 45063UQ12 may contain a value sheet for a different HbA1c calibrator lot. If calibrator lot 45063UQ12 is used in the calibration of the HbA1c assay using values from an incorrect calibrator value sheet then all results, including quality control and patient, may be falsely elevated.
Identifiers
| code_info | List Number: 4P52-02; Lot Number: 45063UQ12; Expiration Date: 15 DEC 2014 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0135-2015%22