RecallRadar

FDA Device recall Z-0133-2023

da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision System…

On 2022-11-02 the FDA classified a Class II recall of da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision System… by Intuitive Surgical, citing inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0133-2023
ClassificationClass II
Statusongoing
Initiated2022-09-20
Published2022-11-02
Last updated2026-09-13 11:56 UTC
CompanyIntuitive Surgical
DistributionAK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MO, MS, NC, NE, NV, OH, PA, SC, TN, TX, WA, WI

Product

da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision System (VSS)

Reason

Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery

Identifiers

code_info
Software version P10 (OS4 v10.0.0/A70_P10_B738) SOFTWARE,EMBEDDED RLS,IS4000 AND IS4200,A70_P10_B738 (Model: 610092-7…Software version P10 (OS4 v10.0.0/A70_P10_B738) SOFTWARE,EMBEDDED RLS,IS4000 AND IS4200,A70_P10_B738 (Model: 610092-738) on Model Name/Model Number/UDI-DI: ASSY,PSCART,IS4000,4-ARM,P10/380652-55/00886874110720, ASSY,SSC,IS4000,P10/380677-28/00886874110744, ASSY,VSS VISION SYSTEM,IS4000,P10/381121-45/00886874110898, ASSY,PSCART,IS4200,P10/380620-47/00886874115404, ASSY,VSC,IS4000,RECONDITIONED,P10/380721-15/00886874115343, ASSY,SSC,IS4000,RECONDITIONED,P10/380723-16/00886874115374

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0133-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.