FDA Device recall Z-0132-2026
The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field u…
- FDA Device
- Class II
- ongoing
- Published 2025-10-22
On 2025-10-22 the FDA classified a Class II recall of The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field u… by Insulet, citing the failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhon….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0132-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-09-11 |
| Published | 2025-10-22 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Insulet |
| Distribution | US |
Product
The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.
Reason
The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.
Identifiers
| code_info | Omnipod 5 App Software. UDI: 10385083000459. All software versions 2.X.X |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.