RecallRadar

FDA Device recall Z-0127-2013

Perfusion System 8000, base 5 pump, 115V safety monitor included The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 h…

On 2012-10-31 the FDA classified a Class II recall of Perfusion System 8000, base 5 pump, 115V safety monitor included The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 h… by Terumo Cardiovascular Systems, citing terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the power distribution and battery back-up systems for the Sarns Mod….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0127-2013
ClassificationClass II
Statusterminated
Initiated2012-09-25
Published2012-10-31
Last updated2026-09-13 11:53 UTC
CompanyTerumo Cardiovascular Systems
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, OR, PA, PR, SC, TN, TX, VA, WA, WI, WV, WY

Product

Perfusion System 8000, base 5 pump, 115V safety monitor included The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Reason

Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the power distribution and battery back-up systems for the Sarns Modular Perfusion System 8000. Very few of the reports occurred during cardiopulmonary bypass; the majority of the malfunctions occurred during set-up or were reported during routine preventative maintenance. The malfunction

Identifiers

code_infoCatalog number: 16401 and serial numbers: 1001-1243, 1247-1648, 1650-1821, and 1824-3003.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0127-2013%22