FDA Device recall Z-0123-2013
UniCel¿ DxC 600/800 with Software v5.0.11, Part Number B12218
- FDA Device
- Class II
- terminated
- Published 2012-10-31
On 2012-10-31 the FDA classified a Class II recall of UniCel¿ DxC 600/800 with Software v5.0.11, Part Number B12218 by Beckman Coulter, citing beckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment procedure cannot be performed on certain DxC configurations with v5.0.11 software instal….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0123-2013 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-09-03 |
| Published | 2012-10-31 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Beckman Coulter |
| Distribution | OH, TX |
Product
UniCel¿ DxC 600/800 with Software v5.0.11, Part Number B12218. The UniCel¿ DxC Synchron¿ Clinical Systems are fully automated, computercontrolled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).
Reason
Beckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment procedure cannot be performed on certain DxC configurations with v5.0.11 software installed. If the Shuttle to Barcode Reader becomes misaligned, the Operator will not be able to perform this alignment procedure on the DxC systems listed above. Software v5.0 .11 does not allow access to the Shuttle to Bar
Identifiers
| code_info | Part Number B12218. Serial numbers 2329, 5477, 5479. |
|---|