FDA Device recall Z-0121-2020
Centurion CIRCLAMP W/1.3 CM BELL-single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male…
- FDA Device
- Class II
- terminated
- Published 2019-10-16
On 2019-10-16 the FDA classified a Class II recall of Centurion CIRCLAMP W/1.3 CM BELL-single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male… by Centurion Medical Products, citing A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0121-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-08-28 |
| Published | 2019-10-16 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Centurion Medical Products |
| Distribution | US |
Product
Centurion CIRCLAMP W/1.3 CM BELL-single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 330CR
Reason
A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.
Identifiers
| code_info | Batch/Lot: 2019061201 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0121-2020%22