FDA Device recall Z-0118-2014

Centricity PACS IW by GE Healthcare Dynamic Imaging Solutions is a device that receives medical images (including mammograms) and data from…

On 2013-11-06 the FDA classified a Class II recall of Centricity PACS IW by GE Healthcare Dynamic Imaging Solutions is a device that receives medical images (including mammograms) and data from… by Ge Healthcare It, citing multiple issues were identified with the GE Centricity PACS-IW system that could pose a safety risk to patients. These issues include: 1) With Auto-Fetch in use another patient's….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0118-2014
ClassificationClass II
Statusterminated
Initiated2012-06-15
Published2013-11-06
Last updated2026-09-13 11:53 UTC
CompanyGe Healthcare It
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY

Product

Centricity PACS IW by GE Healthcare Dynamic Imaging Solutions is a device that receives medical images (including mammograms) and data from various imaging sources. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images may only be interpreted using an FDA approved monitor that offers at least 5 M pixel resolution and meets other technical specifications reviewed and accepted by the FDA. Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians an assistants.

Reason

Multiple issues were identified with the GE Centricity PACS-IW system that could pose a safety risk to patients. These issues include: 1) With Auto-Fetch in use another patient's image may appear without user's intention. 2.a) In the case where dose administration time and image acquisition time fall on either side of midnight, the value will be under reported and close to zero for all pixe

Identifiers

code_infoSoftware Versions: 3.7.1, 3.7.1.1, 3.7.2, 3.7.3.1, 3.7.3.3, 3.7.3.4, 3.7.3.5, 3.7.6.3, 3.7.3.7, and 3.7.3.8

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.