FDA Device recall Z-0116-2025
Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO MOB UPGRADE US
- FDA Device
- Class II
- ongoing
- Published 2024-10-23
On 2024-10-23 the FDA classified a Class II recall of Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO MOB UPGRADE US by Boston Scientific, citing boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0116-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-09-05 |
| Published | 2024-10-23 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Boston Scientific |
| Distribution | US |
Product
Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO MOB UPGRADE US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includes intravascular ultrasound (IVUS). When used with a peripheral vascular imaging catheter, the system utilizes the IVUS modality for ultrasound examination of intravascular pathology.
Reason
Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
Identifiers
| code_info | Material Number H7492493121C0, GTIN/UDI/DI 00191506033194, Batch/Serial Numbers: 102946988, 102946990, 102949556, 1031…Material Number H7492493121C0, GTIN/UDI/DI 00191506033194, Batch/Serial Numbers: 102946988, 102946990, 102949556, 103199497, 103199500, 103199502, 103207815, 103207817, 103207860, 103439257, 103439962, 104233565, 104233569, 104233592, 104260813, 104278173, 104284141, 104322013, 104322062, 104322063, 104322064, 104322065, 104329117, 104329122, 104330747, 104330955, 104333933, 104340756, 104340852, 104340854, 104341075, 104350753, 104350860, 104350861, 104599296, 104599354, 104601145, 104601146, 104601151, 104601162 |
|---|