FDA Device recall Z-0109-2022
Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up…
- FDA Device
- Class II
- ongoing
- Published 2021-10-20
On 2021-10-20 the FDA classified a Class II recall of Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up… by Draeger Medical, citing software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 sec….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0109-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2021-09-10 |
| Published | 2021-10-20 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Draeger Medical |
| Distribution | US |
Product
Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight. Catalog Number: 8417400
Reason
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds
Identifiers
| code_info | Software 2.51 and Lower |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0109-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.