Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED35…

On 2025-10-22 the FDA classified a Class II recall of Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED35… by Microvention, citing flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly a….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0108-2026
ClassificationClass II
Statusongoing
Initiated2025-09-05
Published2025-10-22
Last updated2026-09-13 11:57 UTC
CompanyMicrovention
DistributionUS

Product

Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED3511, FRED5014, FRED3516, MV-F451827; FRED X 27, REF: XFRED4017-MVE, XFRED4528-MVE, MV-F501427X, XFRED4518-MVE, XFRED5526-MVE, XFRED4539-MVE, FREDX5522-PMA, FREDX3522-PMA, FREDX4525-PMA, FREDX5015-PMA, FREDX5532-PMA, FREDX4013-PMA, FREDX3513-PMA, FREDX4515-PMA, FREDX3517-PMA, MV-F352427X

Reason

Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences.

Identifiers

code_info
REF/UDI-DI(Lot): FRED5019/00811425020388(0000235839), 00842429117163(0000544498); FRED5029/00811425020395(0000256343),…REF/UDI-DI(Lot): FRED5019/00811425020388(0000235839), 00842429117163(0000544498); FRED5029/00811425020395(0000256343), 00842429117170(0000507209); FRED4528/00811425020425(0000265666); FRED4026/00811425021033(0000333125, 0000576973); FRED5526/00842429117194(0000361117); FRED4508/00842429117095(0000372088); FRED5514/00842429117187(0000429786, 0000478385); FRED3536/00842429117033(0000460268, 0000445555); FRED3511/00842429117002(0000452923); FRED5014/00842429117156(0000563591, 0000464401); FRED3516/00842429117019(0000513146, 0000531733, 0000536217); MV-F451827/04987892121906(0000565094); XFRED4017-MVE/00842429107508(0000478838); XFRED4528-MVE/00842429107560(0000518323, 0000663751); MV-F501427X/04987892151231(0000579031); XFRED4518-MVE/00842429107553(0000663074, 0000663750); XFRED5526-MVE/00842429107638(0000704203); XFRED4539-MVE/00842429107577(00007311620); FREDX5522-PMA/00842429114483(0000265641, 0000469609, 0000731155); FREDX3522-PMA/00842429114315(0000469598); FREDX4525-PMA/00842429