FDA Device recall Z-0106-2023
CENTURION SnagFree Chest Tube insertion Tray, Reorder CHT545
- FDA Device
- Class II
- terminated
- Published 2022-10-26
On 2022-10-26 the FDA classified a Class II recall of CENTURION SnagFree Chest Tube insertion Tray, Reorder CHT545 by Medline Industries Northfield, citing one lot of chest tube insertion trays, containing a silk suture component that is restricted to 1 cycle of sterilization, was processed though 2 cycles.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0106-2023 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2022-09-07 |
| Published | 2022-10-26 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Medline Industries Northfield |
| Distribution | US |
Product
CENTURION SnagFree Chest Tube insertion Tray, Reorder CHT545
Reason
One lot of chest tube insertion trays, containing a silk suture component that is restricted to 1 cycle of sterilization, was processed though 2 cycles.
Identifiers
| code_info | UDI/DI 20653160222979, Lot Code 2022050690 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0106-2023%22