RecallRadar

FDA Device recall Z-0106-2020

TRIPLE LUMEN CVC 20CM BUNDLE STCVC2040

On 2019-10-16 the FDA classified a Class II recall of TRIPLE LUMEN CVC 20CM BUNDLE STCVC2040 by Centurion Medical Products, citing firm distributes kits which include the BD MaxZeroNeedleess Connector, which was recalled by the manufacturer due to the potential for droplets of fluid separating from the sealed….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0106-2020
ClassificationClass II
Statusterminated
Initiated2019-04-24
Published2019-10-16
Last updated2026-09-13 11:55 UTC
CompanyCenturion Medical Products
DistributionUS

Product

TRIPLE LUMEN CVC 20CM BUNDLE STCVC2040

Reason

Firm distributes kits which include the BD MaxZeroNeedleess Connector, which was recalled by the manufacturer due to the potential for droplets of fluid separating from the sealed surface of the connector following disconnection.

Identifiers

code_info2018110150 2018120350 2018121050 2019021150 2019021550

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0106-2020%22