Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping System and PediMag Blood Pumping System

On 2018-10-24 the FDA classified a Class II recall of Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping System and PediMag Blood Pumping System by Thoratec Switzerland, citing reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0103-2019
ClassificationClass II
Statusterminated
Initiated2018-09-05
Published2018-10-24
Last updated2026-09-13 11:54 UTC
CompanyThoratec Switzerland
DistributionAL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV

Product

Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping System and PediMag Blood Pumping System.

Reason

Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection.

Identifiers

code_infoUDI: 07640135140078. All serial numbers until design change is implemented to make the motor cable more resistant to damage.