FDA Device recall Z-0100-2017
Roche COBAS INTEGRA 800 /800 CTS Analyzer Analyzer, Chemistry (Photometric, Discrete for Clinical use Product Usage: The system cobas p612…
- FDA Device
- Class II
- terminated
- Published 2016-11-02
On 2016-11-02 the FDA classified a Class II recall of Roche COBAS INTEGRA 800 /800 CTS Analyzer Analyzer, Chemistry (Photometric, Discrete for Clinical use Product Usage: The system cobas p612… by Roche Diagnostics Operations, citing roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0100-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-07-15 |
| Published | 2016-11-02 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Roche Diagnostics Operations |
| Distribution | US |
Product
Roche COBAS INTEGRA 800 /800 CTS Analyzer Analyzer, Chemistry (Photometric, Discrete for Clinical use Product Usage: The system cobas p612 is a computer-controlled stand-alone system for aliquoting and sorting of barcoded sample tubes. The system operates with an online connection to the host.
Reason
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Identifiers
| code_info | Not applicable |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0100-2017%22