RecallRadar

FDA Device recall Z-0098-2017

Modular Pre Analytics ( EVO/Plus) Product Usage: The RSD 800A is a computer-controlled stand-online system for decapping, aliquoting and so…

On 2016-11-02 the FDA classified a Class II recall of Modular Pre Analytics ( EVO/Plus) Product Usage: The RSD 800A is a computer-controlled stand-online system for decapping, aliquoting and so… by Roche Diagnostics Operations, citing roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0098-2017
ClassificationClass II
Statusterminated
Initiated2016-07-15
Published2016-11-02
Last updated2026-09-13 11:54 UTC
CompanyRoche Diagnostics Operations
DistributionUS

Product

Modular Pre Analytics ( EVO/Plus) Product Usage: The RSD 800A is a computer-controlled stand-online system for decapping, aliquoting and sorting of primary tubes and for producing and sorting bar coded secondary tubes.. The system operates with an online connection to the host.

Reason

Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

Identifiers

code_infoNot applicable

Status history