FDA Device recall Z-0097-2017
cobas 8000 Modular Analyzer Series c8000, Chemistry (Photometric, Discrete) for clinical use Modular e602: Immunoassay Analyzer Product Usa…
- FDA Device
- Class II
- terminated
- Published 2016-11-02
On 2016-11-02 the FDA classified a Class II recall of cobas 8000 Modular Analyzer Series c8000, Chemistry (Photometric, Discrete) for clinical use Modular e602: Immunoassay Analyzer Product Usa… by Roche Diagnostics Operations, citing roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0097-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-07-15 |
| Published | 2016-11-02 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Roche Diagnostics Operations |
| Distribution | US |
Product
cobas 8000 Modular Analyzer Series c8000, Chemistry (Photometric, Discrete) for clinical use Modular e602: Immunoassay Analyzer Product Usage: The VS II system is a stand-alone, computer-controlled instrument designed to aliquot patient sample from a primary tube to a barcoded secondary tube for processing on an off-line analyzer.
Reason
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Identifiers
| code_info | Not applicable |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0097-2017%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.