FDA Device recall Z-0095-2015
The GemStar Power Supply is an accessory for the GemStar Infusion Pump
- FDA Device
- Class I
- terminated
- Published 2014-11-05
On 2014-11-05 the FDA classified a Class I recall of The GemStar Power Supply is an accessory for the GemStar Infusion Pump by Hospira, citing the GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0095-2015 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2014-02-03 |
| Published | 2014-11-05 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Hospira |
| Distribution | AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY |
Product
The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC.
Reason
The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.
Identifiers
| code_info | Wall Mount GemStar Power Supply List Number: 13026; Lot Numbers: 25283G1, 26286G1, 26287G1, 27292G1, 27294G1, 27295G1, 27296G1 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0095-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.