FDA Device recall Z-0094-2023
N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299
- FDA Device
- Class II
- ongoing
- Published 2022-10-26
On 2022-10-26 the FDA classified a Class II recall of N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299 by Siemens Healthcare Diagnostics, citing N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0094-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-08-30 |
| Published | 2022-10-26 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299
Reason
N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).
Identifiers
| code_info | UDI-DI: 0405686900179VF Lot Number: 153083; 153085; 153087A; 153090; 153095B; 153099A; 153004B. Future lots may be i…UDI-DI: 0405686900179VF Lot Number: 153083; 153085; 153087A; 153090; 153095B; 153099A; 153004B. Future lots may be impacted by the issue but will then be provided with a corresponding instruction note within the reagent package |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0094-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.