FDA Device recall Z-0093-2025

Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device

On 2024-10-23 the FDA classified a Class III recall of Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device by Beckman Coulter, citing beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results wh….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0093-2025
ClassificationClass III
Statusongoing
Initiated2024-07-16
Published2024-10-23
Last updated2026-09-13 11:57 UTC
CompanyBeckman Coulter
Distributionn/a

Product

Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.

Reason

Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with iChemVELOCITY Urine Chemistry Strips (PNs 800-7204 and 800-7212) on the iChemVELOCITY Analyzer. This issue may delay the reporting of patient results, with a worst-case scenario of a delay in diagnosis or treatment of metabolic disorders, kidney function abnormalities, urinary tract infections, and liver function.

Identifiers

code_infoUDI/DI 10837461002611, Lot Numbers: 432-24 and higher

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.