Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097

On 2021-10-27 the FDA classified a Class I recall of Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097 by Datascope, citing A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming b….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0091-2022
ClassificationClass I
Statusongoing
Initiated2021-09-21
Published2021-10-27
Last updated2026-09-13 11:56 UTC
CompanyDatascope
DistributionUS

Product

Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097. Component of the Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) with the below Model Numbers. 1. Cardiosave Hybrid. Model Numbers: 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, 0998-UC-0800-55. 2. Cardiosave Rescue. Model Number: 0998-00-0800-83.

Reason

A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to customers.

Identifiers

code_info
Battery Pack Serial Numbers: 192224349IP, 191908720IP, 202369811IP, 202791944IP, 191848115IP, 192099841IP, 192103741IP,…Battery Pack Serial Numbers: 192224349IP, 191908720IP, 202369811IP, 202791944IP, 191848115IP, 192099841IP, 192103741IP, 192104141IP, 171218521PE, 181642339PE, 202321308IP, 191912520IP, 191874716IP, 191922620IP, 181642639PE, 192185446IP, 171277935PE, 192190247IP, 202400219IP, 202323808IP, 191910720IP, 181661142PE, 192245250IP, 202599332IP, 181645841PE, 181526530PE, 202403719IP, 181705244PE, 192104541IP, 191787312IP, 181646041PE, 191955422IP, 202389919IP, 191822513IP, 191853415IP, 192004724IP, 192196447IP, 192155645IP, 202330609IP, 171325749PE, 192018828IP, 191778911IP, 192187947IP, 192196247IP, 181609237PE, 192098341IP, 191972823IP, 191860615IP, 202719740IP, 202334109IP, 202651035IP, 191852615IP, 192173146IP, 202386519IP, 202398619IP, 202416720IP, 202599432IP, 191956322IP, 192205847IP, 192220349IP, 192081138IP, 202617233IP, 181659242PE, 202266705IP, 171255732PE, 192179046IP, 192179246IP, 202280605IP, 181673143PE, 192083838IP, 192106442IP, 192216649IP, 191956122IP, 181721730IP, 191943120