FDA Device recall Z-0091-2015

Codman Lumbar Drainage Catheter Kit II with EDS 3 (External Drainage System) Catalog Number: 82-1738 Indicated for draining cerebrospinal f…

On 2014-10-29 the FDA classified a Class II recall of Codman Lumbar Drainage Catheter Kit II with EDS 3 (External Drainage System) Catalog Number: 82-1738 Indicated for draining cerebrospinal f… by Codman And Shurtleff, citing tubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or under-drainage of CSF from the ventricular system or introduction of air….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0091-2015
ClassificationClass II
Statusterminated
Initiated2014-09-15
Published2014-10-29
Last updated2026-09-13 11:53 UTC
CompanyCodman And Shurtleff
DistributionUS

Product

Codman Lumbar Drainage Catheter Kit II with EDS 3 (External Drainage System) Catalog Number: 82-1738 Indicated for draining cerebrospinal fluid (CFS) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated

Reason

tubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or under-drainage of CSF from the ventricular system or introduction of air into the ventricular system (pneumocephalus)

Identifiers

code_infoAll lots with expiration on or before August 2017 (2017-08)