RecallRadar

FDA Device recall Z-0086-2017

15 g R.K. Epidural Needle (TW), Catalog #/REF #100-1415 Intended for the epidural placement, directly or through an epidural catheter, of a…

On 2016-11-09 the FDA classified a Class II recall of 15 g R.K. Epidural Needle (TW), Catalog #/REF #100-1415 Intended for the epidural placement, directly or through an epidural catheter, of a… by Epimed International, citing there is a possibility that the needles may have the potential to skive a catheters outer coating. If this were to occur, there exists the risk that a portion of the catheter cou….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0086-2017
ClassificationClass II
Statusterminated
Initiated2016-04-22
Published2016-11-09
Last updated2026-09-13 11:54 UTC
CompanyEpimed International
DistributionAK, AL, FL, IN, TN, TX

Product

15 g R.K. Epidural Needle (TW), Catalog #/REF #100-1415 Intended for the epidural placement, directly or through an epidural catheter, of anesthetic agents to elicit regional anesthesia

Reason

There is a possibility that the needles may have the potential to skive a catheters outer coating. If this were to occur, there exists the risk that a portion of the catheter could remain in the patient, which may require otherwise unnecessary medical invention.

Identifiers

code_info
1) Spectra Lot# GC091702 (Epimed manufacturing lot # 2801501, Sterile Lot # 12157181) 2) Spectra Lot# GC101871 (Epimed…1) Spectra Lot# GC091702 (Epimed manufacturing lot # 2801501, Sterile Lot # 12157181) 2) Spectra Lot# GC101871 (Epimed manufacturing lot # 3211503, Sterile Lot # 12157445) 3) Spectra Lot# GC101897 (Epimed manufacturing lot # 3161501, Sterile Lot # 12157325)

Status history