FDA Device recall Z-0086-2015

Siemens Mammomat Inspiration mammography systems

On 2014-10-29 the FDA classified a Class II recall of Siemens Mammomat Inspiration mammography systems by Siemens Medical Solutions Usa, citing the stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0086-2015
ClassificationClass II
Statusterminated
Initiated2014-09-18
Published2014-10-29
Last updated2026-09-13 11:53 UTC
CompanySiemens Medical Solutions Usa
DistributionMA, NY, VA

Product

Siemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

Reason

The stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already i

Identifiers

code_infomaterial #10140000, with serial numbers: 5018 3849 5042

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.