Siemens Mammomat Inspiration mammography systems
- FDA Device
- Class II
- terminated
- Published 2014-10-29
On 2014-10-29 the FDA classified a Class II recall of Siemens Mammomat Inspiration mammography systems by Siemens Medical Solutions Usa, citing the stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in th….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0086-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-09-18 |
| Published | 2014-10-29 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | MA, NY, VA |
Product
Siemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.
Reason
The stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already i
Identifiers
| code_info | material #10140000, with serial numbers: 5018 3849 5042 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0086-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.