RecallRadar

FDA Device recall Z-0084-2025

Auxiliary Common Gas Outlet (ACGO) conversion/install kits sold under product names CONV KIT SCGO TO ACGO and ACGO ASSY FIELD INSTL. These…

On 2024-10-30 the FDA classified a Class I recall of Auxiliary Common Gas Outlet (ACGO) conversion/install kits sold under product names CONV KIT SCGO TO ACGO and ACGO ASSY FIELD INSTL. These… by Datex Ohmeda, citing potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0084-2025
ClassificationClass I
Statusongoing
Initiated2024-07-12
Published2024-10-30
Last updated2026-09-13 11:57 UTC
CompanyDatex Ohmeda
DistributionUS

Product

Auxiliary Common Gas Outlet (ACGO) conversion/install kits sold under product names CONV KIT SCGO TO ACGO and ACGO ASSY FIELD INSTL. These kits are not affected specifically, however are included due to being purchased with or later used to convert to ACGO.

Reason

Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

Identifiers

code_info
No GTIN. Spare Part Numbers: 1009-8249-000, 1011-8071-000. Not model or lot specific. All devices of the indicated pr…No GTIN. Spare Part Numbers: 1009-8249-000, 1011-8071-000. Not model or lot specific. All devices of the indicated products configured with the AGCO option are potentially affected.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0084-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.