FDA Device recall Z-0084-2017
Phoenix Recon Targeting Arm Guide, Surgical, Instrument for the sleeve and Orthopedic Manual Surgical Instruments for the Arm Product Usage…
- FDA Device
- Class II
- terminated
- Published 2016-10-19
On 2016-10-19 the FDA classified a Class II recall of Phoenix Recon Targeting Arm Guide, Surgical, Instrument for the sleeve and Orthopedic Manual Surgical Instruments for the Arm Product Usage… by Zimmer Biomet, citing old revisions and new revisions of the soft tissue sleeve and recon targeting arm are not interchangeable, as the new revision tissue sleeve will not fit into the old revision tar….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0084-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-08-09 |
| Published | 2016-10-19 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Zimmer Biomet |
| Distribution | US |
Product
Phoenix Recon Targeting Arm Guide, Surgical, Instrument for the sleeve and Orthopedic Manual Surgical Instruments for the Arm Product Usage: The Phoenix Femoral Nail System is to be implanted into the femur for alignment, stabilization and fixation of fractures caused by trauma or disease, and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity, and for arthrodesis. These instruments are used in support of the surgery.
Reason
Old revisions and new revisions of the soft tissue sleeve and recon targeting arm are not interchangeable, as the new revision tissue sleeve will not fit into the old revision targeting arm. The old revision tissue sleeve has too much clearance with new revision targeting arm. Risks of encountering the affected product include: the correct instrument may not be readily available for the procedure causing the surgeon to complete the procedure by hand; or a delay in surgery greater than 30 minutes may occur while another Recon Targeting Arm or Recon Soft tissue Sleeve is located or while the surgeon completes the procedure by hand.
Identifiers
| code_info | Item number: 14-442018 Lot number: 290610 290620 386780 |
|---|