FDA Device recall Z-0080-2017

Zero-Gravity Radiation Protection System Radiology Personnel protective shield

On 2016-10-19 the FDA classified a Class II recall of Zero-Gravity Radiation Protection System Radiology Personnel protective shield by Contour Fabricators, citing two customer complaints suggest that excessive rotation of the cable in one direction (clockwise or counter-clockwise) without allowing the cable to unwind to its normal condition….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0080-2017
ClassificationClass II
Statusterminated
Initiated2016-09-09
Published2016-10-19
Last updated2026-09-13 11:54 UTC
CompanyContour Fabricators
DistributionUS

Product

Zero-Gravity Radiation Protection System Radiology Personnel protective shield.

Reason

Two customer complaints suggest that excessive rotation of the cable in one direction (clockwise or counter-clockwise) without allowing the cable to unwind to its normal condition may cause the cable to break suddenly and the body shield to fall. While no injuries or deaths occurred in either failure, the Zero-GravityTM body shield weighs 55 pounds, and in the event of a cable failure, the body shield will fall immediately. A fall of this nature places the user, associated staff and the patient at risk for a potential injury.

Identifiers

code_info
All serial numbers for all models listed below manufactured/distributed between June 28, 2010  August 25, 2016. Par…All serial numbers for all models listed below manufactured/distributed between June 28, 2010  August 25, 2016. Part No. : Description ZGM-6-.5H : Zero-Gravity floor unit ZGM-6-5ROUNDH : Zero-Gravity floor unit ZGM-6Demo : Zero-Gravity demo floor unit ZGCM-48 : Zero-Gravity ceiling mounted monorail with 48 boom arm ZGCM-48-SPECIAL : Zero-Gravity ceiling mounted monorail with 48 boom arm with monorail extension ZG48 : Zero-Gravity 48 boom arm assembly and body shield for monorail ZGCM-66 : Zero-Gravity ceiling mounted monorail with 66 boom arm ZGCM-HSA : Zero-Gravity ceiling mounted monorail with hinged swing arm ZGCH : Zero-Gravity ceiling mounted monorail Contour track with 48 boom arm ZGCH48 : Zero-Gravity 48 boom arm assembly and body shield for Contour track ZGHSA : Zero-Gravity single point, ceiling mount with hinged swing arm ZGCSP : Zero-Gravity single point, ceiling mount with hinged swing arm 29858 : Zero-Gravity floor unit