FDA Device recall Z-0079-2019
VITROS 5600 Integrated System-For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes…
- FDA Device
- Class II
- terminated
- Published 2018-10-17
On 2018-10-17 the FDA classified a Class II recall of VITROS 5600 Integrated System-For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes… by Ortho Clinical Diagnostics, citing potential for the VITROS 5600 Integrated System to dispense an insufficient sample volume onto a MicroSlide without suppressing the potentially incorrect result.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0079-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-08-15 |
| Published | 2018-10-17 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Ortho Clinical Diagnostics |
| Distribution | US |
Product
VITROS 5600 Integrated System-For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, Catalog Number: 6802413
Reason
Potential for the VITROS 5600 Integrated System to dispense an insufficient sample volume onto a MicroSlide without suppressing the potentially incorrect result.
Identifiers
| code_info | Serial Numbers : J56000024 through J56003300 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0079-2019%22