RecallRadar

FDA Device recall Z-0074-2027

VERIFY Biological Steam Test Pack Model/Catalog Number: S3065

On 2026-10-14 the FDA classified a Class II recall of VERIFY Biological Steam Test Pack Model/Catalog Number: S3065 by Steris, citing geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potentia….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0074-2027
ClassificationClass II
Statusongoing
Initiated2026-08-31
Published2026-10-14
Last updated2026-10-21 05:30 UTC
CompanySteris
DistributionUS

Product

VERIFY Biological Steam Test Pack Model/Catalog Number: S3065

Reason

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Identifiers

code_info
UDI: 10724995220300 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: exp…UDI: 10724995220300 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-10-21 05:30 UTC. Current status: ongoing.

Other recalls by Steris

All Steris recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0074-2027%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.