RecallRadar

FDA Device recall Z-0074-2025

Aespire 7100, 7100 (Variant), and Aespire 7100 S/5 Protiva configurations

On 2024-10-30 the FDA classified a Class I recall of Aespire 7100, 7100 (Variant), and Aespire 7100 S/5 Protiva configurations by Datex Ohmeda, citing potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0074-2025
ClassificationClass I
Statusongoing
Initiated2024-07-12
Published2024-10-30
Last updated2026-09-13 11:57 UTC
CompanyDatex Ohmeda
DistributionUS

Product

Aespire 7100, 7100 (Variant), and Aespire 7100 S/5 Protiva configurations. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.

Reason

Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.

Identifiers

code_info
a. Aespire 7100 - GTIN: 00840682102261. Model Number: 1009-9000-000. All Serial Numbers. b. Aespire 7100 (Variant) -…a. Aespire 7100 - GTIN: 00840682102261. Model Number: 1009-9000-000. All Serial Numbers. b. Aespire 7100 (Variant) - GTIN: 00840682102261. Model Numbers: 1009-9101-000 and 1009-9101-000. All Serial Numbers. c. Aespire 7100 S/5 Protiva - No GTIN. Model Number: 1009-9004-000. All Serial Numbers. Not model or lot specific. All devices of the indicated products configured with the AGCO option are potentially affected.