Electrohydraulic lithotripter probe, 1.9 FR 375 cm, Model 9-195-371DS used with a Model 9-201-00 Autolith lithotripter. electrohydraulic li…

On 2017-11-15 the FDA classified a Class II recall of Electrohydraulic lithotripter probe, 1.9 FR 375 cm, Model 9-195-371DS used with a Model 9-201-00 Autolith lithotripter. electrohydraulic li… by Northgate Technologies, citing the inner and outer labels on the product contain incorrect reference number of 9-202-3751DS. According to Northgate Technologies, the resistor in the probe will not allow the mis….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0073-2018
ClassificationClass II
Statusterminated
Initiated2017-05-24
Published2017-11-15
Last updated2026-09-13 11:54 UTC
CompanyNorthgate Technologies
DistributionMA, NE, TN

Product

Electrohydraulic lithotripter probe, 1.9 FR 375 cm, Model 9-195-371DS used with a Model 9-201-00 Autolith lithotripter. electrohydraulic lithotripsy (EHL) is a method that allows large stones to be fragmented and removed percutaneously or endoscopically.

Reason

The inner and outer labels on the product contain incorrect reference number of 9-202-3751DS. According to Northgate Technologies, the resistor in the probe will not allow the mislabeled product to be used with the incorrect generator. Could Delay Care.

Identifiers

code_infoModel No. 9-195-371DS, Lot/Serial Number LDS12516

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0073-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.